When the camera is used in high radiation environments, the major gamma emission may be outside the FOV of the camera. In this case it is important that:
Compton cameras are superior to collimator-based systems in such situations because they can detect imaging radiation on 4π. On the Compton cameras, our camera is specially suited for those situations because:
Below are tests illustrating our claims:
1st experiment – no source at 0° + 0.7MBq 22Na source at 180° – We observe the 22Na source behind the camera and some noise in front of the camera.
This experiment proves the exceptional sensitivity of the Temporal camera.
2nd experiment – a weak 0.03MBq 137Cs source at 0° + 0.7MBq 22Na source at 180° – We obtain a perfect image despite the strong perturbation.
3rd experiment – a weak 0.1MBq 22Na source at 0° + 0.7MBq 22Na source at 180° – We obtain the image with some noise, but the source is correctly reconstructed.
This experiment proves the capability of Temporal camera to observe weak sources (>1:10 contrast) in the 0° sector even with strong perturbations behind.